Power Harmonics and Leakage Tester Prova19/21/23
Quantity
Product Features
Model Prova 19, 21, 23
- True RMS value (V and A)
- Leakage Current Measurement at 10 μ A Resolution
- Harmonic Analysis (V and A) to the 99 th Order
- Better Understanding of High Frequency Harmonic Analysis (up to 5/6 KHz)
- Non-interrupted Harmonic Analysis
- Analysis of Total Harmonic Distortion (THD-F)
- Analysis of Crest Factor (C.F.)
- Fast Peak Function (33 μ s and 39 μ s)
- Programmable CT Ratio (1 to 250)
- W, KW, H.P.
- VA, KVA, VAR, KVAR
- PF, Phase Angle ( Φ )
- Energy (mWH, WH, KWH)
- Analysis of Standby Power Consumption
- Balanced 3 Φ Power Quality
- Balanced 3 Φ Sequence
Please refer to users manual for more detailed and up-to-date description of specification. |
AC Watt (50 or 60 Hz, PF 0.5 to 1. CT = 1, Accuracy of Readings) Model 21 (0 to 50A), Model 23 (0 to 100A) |
Range
|
Resolution
|
Accuracy 21 (0 - 30A) Accuracy 23 (0 - 60A) |
Accuracy 21 (30 - 50A) Accuracy 23 (60 - 100A) |
0.000 – 9.999 W
|
0.001 W
|
±1% ± 0.02 W
|
±2% of VA ± 5dgts
|
10.00 – 99.99 W
|
0.01 W
|
±1% ± 0.2 W
|
±2% of VA ± 5dgts
|
100.0 – 999.9 W
|
0.1 W |
±1% ± 0.2 W
|
±2% of VA ± 5dgts
|
1.000 – 9.999 KW
|
0.001 KW |
±1% ± 0.02 KW
|
±2% of VA ± 5dgts
|
10.00 – 99.99 KW
|
0.01 KW |
±1% ± 0.2 KW
|
±2% of VA ± 5dgts
|
100.0 – 999.9 KW
|
0.1 KW |
±1% ± 0.2 KW
|
±2% of VA ± 5dgts
|
1000 – 9999 KW
|
1 KW |
±1% ± 2 KW
|
±2% of VA ± 5dgts
|
AC Current (50 or 60 Hz, True RMS) |
Range (Model 19,21)
|
Range (Model 23)
|
Resolution |
Accuracy of Readings |
0.30 – 60.00mA
|
0.30 – 60.00mA
|
0.01mA
|
±0.5% ± 5dgts
|
60.0 – 600.0mA
|
60.0 – 600.0mA
|
0.1mA
|
±0.5% ± 5dgts
|
0.030 – 3.000 A
|
0.030 – 9.999 A
|
0.001 A
|
±0.5% ± 5dgts
|
3.00 – 30.00 A
|
10.00 – 60.00 A
|
0.01 A
|
±0.5% ± 5dgts
|
30.00 – 50.00 A
|
6 0.00 – 99.99 A
|
0.01 A
|
±1.0% ± 5dgts
|
AC Voltage (50 or 60 Hz, True RMS) |
Range |
Resolution |
Accuracy of R. |
Range |
Resolution |
Accuracy of R. |
3V – 250V
|
0.1 V
|
±0.5% ± 5dgts
|
250V – 600V
|
0.1 V
|
±0.5% ± 5dgts
|
Harmonics of AC Voltage in % and Magnitude (1 to 99th order) |
Range
|
Resolution
|
Accuracy in %
|
Resolution
|
Accuracy in Magnitude
|
50V – 600V
|
0.1 %
|
±0.5%
|
0.1 V
|
±0.5% ± 0.5V
|
Harmonics of ACA in % Percentage and Magnitude (1 to 99th order) |
Range (Model 19, 21) |
Range (Model 23)
|
Res.
|
Accuracy (%)
|
Res.
|
Accuracy of R.
|
10.00 – 60.00mA
|
10.00 – 60.00mA
|
0.1 %
|
±0.5%
|
0.01mA
|
±0.5% ± 0.1mA
|
60.0 – 600.0mA
|
60.0 – 600.0mA
|
0.1 %
|
±0.5%
|
0.1mA
|
±0.5% ± 1mA
|
0.5 – 3.000 A
|
0.5 – 9.999 A
|
0.1 %
|
±0.5%
|
0.001 A
|
±0.5% ± 0.01 A
|
3.00 – 30.00 A
|
10.00 – 60.00 A
|
0.1 %
|
±0.5%
|
0.01 A
|
±0.5% ± 0.1 A
|
30.00 – 60.00 A
|
6 0.00 – 99.99 A
|
0.1 %
|
±1.0%
|
0.01 A
|
±1.0% ± 0.1 A
|
Power Factor and Phase Angle (PF, Φ ,Model 21 and 23) |
Range (PF)
|
Resolution
|
Accuracy
|
Range (Φ)
|
Resolution
|
Accuracy
|
0.000 – 1.000
|
0.001
|
± 1°
|
-180° to 180° (0 ° to 360 ° )
|
0.1°
|
± 1°
|
Total Harmonic Distortion (THD-F, 1 to 99th order) Crest Factor (C.F., Accuracy of Readings) |
Range
|
Resolution
|
Accuracy of R.
|
Range
|
Resolution
|
Accuracy of R.
|
0.0–999.9%
|
0.1%
|
± 0.5% ± 5 digits
|
1.00 – 99.99
|
0.01
|
± 5% ± 30 digits
|
General Specification |
Conductor Size | 30mm (approx.) |
Battery Type | 1.5V SUM-3 x 2 |
Display | 4+2+2 digits LCD |
Power Consumption | 5mA (approx.) |
Auto-Power-Off | 30 minutes |
No. Of Samples/Period | 512 (V & A), 256 (W) |
LCD update rate | 2 times / sec. |
Weight | 200g (battery included) |
Operating Temperature | -10 ° C to 50° C |
Operating Humidity | < 85% relative |
Storage Temperature | -20° C to 60° C |
Storage Humidity | < 75% relative |
Dimension | 210 (L) x 62 (W) x 35.6mm (H), 8.3" (L) x 2.5" (W) x 1.4" (H) |
Accessories | test leads, Carrying bag x 1, Users manual x 1, Batteries 1.5V x 2 |
Option | alligator clips |