The advanced PEN direct reading dosimeters are rugged, precision instruments. They are higher quality and have greater reliability than other dosimeters of this type. They measure and directly read, at any time, accumulated dose (quantity of gamma and x-ray exposure). PEN dosimeters are designed to satisfy military specifications for the RADIAC METER IM-264/PD and ANSI N 13.5 and N322 requirements. A sturdy metal clip attaches the dosimeter to a pocket or any object to monitor total radiation exposure. They are hermetically sealed and immersion proof.

Exposure Range (gamma and x-rays)

PEN200mR Scale increments of 10mR
PEN500mR Scale increments of 20mR
PEN2R increments of 100 mR
PEN5R increments of 200mR
PEN20R increments of 1R
PEN2mSv increments of .1mSvp
PEN5mSv increments of .2mSv
National Stock Numbers Available

Energy Dependence

±10% maximum change in sensitivity for x and gamma ray energies from 16 keV to 6 MeV.

Electrical Leakage

In the absence of radiation, the dosimeters leakage or self discharge is less than 0.5% of full scale in 24 hours at 50° C. At 20° C dosimeters will read background radiation. After gross over exposure totaling up to 2000 R, the leakage is less than 5% of full scale in 48 hours at 20° C. Thereafter, the leakage will decrease to original values.Pen Meter Scale

Charging Voltage

Dosimeters are capable of being charged to "zero" with 140 to 195 volts.

Operational Characteristics

A maximum change in sensitivity of ±10% may occur under any probable combination of the following:
A) Temperature range below -20°C (-4°F) or above +50°C (122°F)
B) Greater than 50,000 foot altitude for one month
C) Relative humidity greater than 95% Highly resistant to shock and vibration.

Accuracy

±10% of true dose for Cs-137 or Co-60 gamma.

Rate Response

Dosimeters are dose rate independent.

Enclosure Material

Barrel - carbon fiber-filled, high-strength, electrically-conducting plastic with metal clip.

Weight

Approx. 19 grams (0.7 oz.)

Size

PEN- 200, 500 2mSv, 5mSv Length: 11 cm (4.48 inches) Diameter: 1.5 cm (0.6 inches) PEN- 2, 5, 20 Length: 11.5 cm (4.52 inches) Diameter: 1.5 cm (0.6 inches)

Options

Protective sapphire lens

Warranty

2 year limited warranty

 

 

Dosi-Lite Pen Dosimeter Light / See Your Readings More Easliy

DosiLiteThe Dosi-Lite provides illumination for a Direct-Reading Dosimeter in low light conditions which is held directly in front of the eye for viewing. The Dosi-Lite comprises of a light-emitting diode (LED) enclosed in a thimble-shaped flexible plastic casing, which slides onto the end of a dosimeter. When the user squeezes the casing, the leads are brought into contact with the batteries, thereby completing the circuit, lighting the LED, and illuminating the Direct-Reading Dosimeter.

The Dosi-Lite is designed for personnel that spend extended periods in areas with lighting which is insufficient or inconvenient for reading the Direct-Reading Dosimeter, thus running the risk of being exposed to excessive radiation before taking the next reading. The Dosi-Lite provides a convenient illuminating device for Direct-Reading Dosimeters that fits onto the end of a Direct-Reading Dosimeter.

Material: Black PVC

Size

0.6" I.D. x 0.71" O.D. x 1.64" Overall Length

Weight

0.5 oz.

Power

(2) A76 Alkaline Battery

Light

LED Ultra Bright Green Water Clear Light

Battery Life

In excess of 800 service hours