ElektroPhysik - Ultrasonic Thickness Gauge - QuintSonic 7
ElektroPhysik - Ultrasonic Thickness Gauge - QuintSonic 7 Coating Thickness Gauges Coating / Paint Testing Malaysia, Selangor, Kuala Lumpur (KL) Supplier, Suppliers, Supply, Supplies | Obsnap Instruments Sdn Bhd
Quantity
Overview

Application

Non-destructive measurement of:

  • Varnish, synthetic materials, glass, enamel and other coatings on wood, synthetic materials, glass, metal and ceramic
  • Metal coatings on synthetic materials, ceramic and wood
  • Up to 5 coatings in a single operation.

Also suited for wall thickness measurement.

Performance features

  • Quick evaluation of the measuring signal, high accuracy and repeatability owing to the sensor-integrated digital 32-bit signal processing  (SIDSP®)
  • Over 40 measuring values per minute: start, wait for 1 second, done.
  • Mere 5 mm measuring area enables gathering measurement on tiny and curved parts
  • Graphic display showing individual coatings, zoom-enabled A screen and menu-driven user guidance

Scope of delivery

  • QuintSonic 7 basic unit
  • Ultrasonic sensor with 1 m / 40' sensor cable
  • USB-Adapter cable
  • Software QSoft 7 Basic Edition on CD
  • Operating manual German / English on CD
  • 1 bottle glycerine couplant, 100 ml
  • 1 bottle coupling gel by ElektroPhysik, 100 ml
  • 4 batteries AA / LR06
  • Plastic carrying case
Technical Details

Property Value
Measuring principle Ultrasonic pulse run-time method
Measuring range /
Resolution /
Accuracy
No. Measuring range Resolution Accuracy
1: 356 µm 0,25 µm 0,25 µm
2: 890 µm 0,5 µm 0,5 µm
3: 1.900 µm 1 µm 1 µm
4: 3.900 µm 2 µm 2 µm
5: 7.500 µm 4 µm 4 µm
(at 2375 m/s ultrasonic velocity in all layers)
Minimum layer thickness approx. 10 µm (depending on the ultrasonic
velocity in the layer)
Number of layers up to 5 layers in one operation
Measurement cycle time approx. 1 s
Measuring area 5 mm Ø / 0.2' Ø
Contact area 10 mm Ø / 0.4' Ø
Minimum curvature radius
convex: 50 mm / 2'
concave: -
No. of test series 300
Measurement storage 250,000
Measurement units µm, mm, mils
Calibration Sound velocity calibration for up to 5 layers
Statistics N (number of measurment values)
Minimum
Maximum
Average
Standard deviation
Coefficient of variation
Block statistics (user adjustable)
Histogram
Trend diagram
Interface Infrared (IrDA® 1.0)
USB
optional RS-232 via additional adapter cable
International Standards DIN EN ISO 2808
Signal processing Sensor integrated digital 32-bit signal processing (SIDSP®)
Operating temperature 5 °C - 50 °C / 40 ° F - 122 °F
Storage temperature -10 °C - 50 °C / 14 °F - 122 °F
Power Supply 4 x Mignon battery AA / LR06
Power supply (90 - 240V AC / 48 - 62 Hz)
Dimensions
Gauge: 153 mm x 89 mm x 32 mm / 6' x 3.5' x 1.3'
Sensor: Ø 25 mm x 60 mm / Ø 1' x 2.4'
Weight
Gauge: 310 g / 10.8 oz
Sensor: 80 g / 2.8 oz
Cable: 30 g / 1 oz

Please leave your enquiry here, we will reply as soon as possible.


*only support gif, jpeg, jpg, png, pdf
Switch To Desktop Version