Hot wall retort furnaces up to 1100ˇăC
Hot wall retort furnaces up to 1100ˇăC Retort Furnaces Nabertherm Furnace Laboratory Equipment Facility Malaysia, Selangor, Kuala Lumpur (KL) Supplier, Suppliers, Supply, Supplies | Obsnap Instruments Sdn Bhd
Quantity
View old post here


These gastight retort furnaces are equipped with direct or indirect heating depending on temperature. They are perfectly suited for various heat treatment processes requiring a defined protective or a reaction gas atmosphere. These compact models can also be laid out for heat treatment under vacuum up to 600 °C. The furnace chamber consists of a gastight retort with water cooling around the door to protect the special sealing. Equipped with the corresponding safety technology, retort furnaces are also suitable for applications under reaction gases, such as hydrogen or, in combination with the IDB package, for inert debinding or for pyrolysis processes. Different model versions are available depending on the temperature range required for the process:

Models NRA ../06 with Tmax 650 °C

  • Heating elements located inside the retort
  • Temperature uniformity up to +/- 5 °C inside the work space
  • Retort made of 1.4571
  • Gas circulation fan in the back of the retort provides for optimal temperature uniformity
  • Insulation made of mineral wool

Models NRA ../09 with Tmax 950 °C

  • Design like models NRA ../06 with following differences:
    • Outside heating with heating elements around the retort
    • Retort made of 1.4828
    • Only fiber materials are used which are not classified as carcinogenic according to TRGS 905, class 1 or 2

Models NR ../11 with Tmax 1100 °C

  • Design like models NRA ../09 with following differences:
    • Retort made of 1.4841
    • Without gas-circulation

 

Standard Equipment for all models

Basic version

  • Compact housing in frame design with removable stainless steel sheets
  • Controls and gas supply integrated in the furnace housing
  • Welded charging supports in the retort resp. air-baffle box in the furnaces with atmosphere circulation
  • Swivel door hinged on right side
  • Open cooling water system
  • Depending on furnace volume for 950 °C- and 1100 °C-models the control system is divided in one or more heating zones
  • Furnace temperature control with measurement outside the retort
  • Gas supply system for one non-flammable protective or reaction gas with flow meter and magnetic valve
  • Port for vacuum pump for cold evacuation
  • Operation under vacuum up to 600 °C with optional single-stage rotary vane pump
  • Defined application within the constraints of the operating instructions
  • NTLog Basic for Nabertherm controller: recording of process data with USB-flash drive

Additional equipment

  • Upgrade for other non-flammable gases, H2 version for flammable gases
  • Automatic gas injection, including MFC flow controller for alternating volume flow, controlled with process control H3700, H1700
  • Vacuum pump for evacuating of the retort up to 600 °C, attainable vacuum up to 10-5 mbar subject to selected pump
  • Indirect cooling
  • Direct cooling
  • Heat exchanger with closed-loop cooling water circuit for door cooling
  • Measuring device for residual oxygen content
  • Door heating
  • Temperature control as charge control with temperature measurement inside and outside the retort
  • Retort, made of 2.4633 for Tmax 1150 °C
  • Process control and documentation via VCD software package or Nabertherm Control Center (NCC) for monitoring, documentation and control

Model Tmax °C Model Tmax °C Working chamber dimensions in mm Useful volume in l Electrical connection
W D H
NRA 17/.. 650 or 950 NR 17/11 1100 225 350 225 17 three-phase
NRA 25/.. 650 or 950 NR 25/11 1100 225 500 225 25 three-phase
NRA 50/.. 650 or 950 NR 50/11 1100 325 475 325 50 three-phase
NRA 75/.. 650 or 950 NR 75/11 1100 325 700 325 75 three-phase
NRA 150/.. 650 or 950 NR 150/11 1100 450 750 450 150 three-phase
NRA 200/.. 650 or 950 NR 200/11 1100 450 1000 450 200 three-phase
NRA 300/.. 650 or 950 NR 300/11 1100 590 900 590 300 three-phase
NRA 400/.. 650 or 950 NR 400/11 1100 590 1250 590 400 three-phase
NRA 500/.. 650 or 950 NR 500/11 1100 720 1000 720 500 three-phase
NRA 700/.. 650 or 950 NR 700/11 1100 720 1350 720 700 three-phase
NRA 1000/.. 650 or 950 NR 1000/11 1100 870 1350 870 1000 three-phase