Mahr Metrology - Desktop Formtester MMQ 400
Quantity
MarForm MMQ 400 is suitable for universal use for extensive workpiece evaluation according to DIN ISO 1101.
High precision measuring axes in Z and X make every form measuring task possible. MarForm MMQ 400 is available in different versions for:
Technical Data
High precision measuring axes in Z and X make every form measuring task possible. MarForm MMQ 400 is available in different versions for:
- High precision workpieces
- Unusually long workpieces
- Large and heavy workpieces
- Use in production or in the precision measuring room
- Motorized or manual centering and tilting table
- Vertical axis (Z) with 900 mm, 500 mm or 350 mm measuring length
- Horizontal axis (X) with 180 mm or 280 mm measuring length and with digital linear scales in the X and Z axes; high probe position accuracy for measurements requiring high reproducibility
- Manual or fully automatic probe with short measuring circuit, high linearity, low measuring force
- Path control for fast measuring of a nominal contour
Technical Data
MMQ 400-2
Z= 350 mm X= 180 mm |
MMQ 400-2
Z= 500 mm X= 280 mm |
MMQ 400-2
Z= 900 mm X= 280 mm |
|
Roundness deviation (µm+µm/mm measuring height)
|
0,01 + 0,00025
|
0,01 + 0,00025
|
0,01 + 0,00025
|
---|---|---|---|
Roundness deviation (µm+µm/mm measuring height)
|
0,02 + 0,0005
|
0,02 + 0,0005
|
0,02 + 0,0005
|
Axial runout deviation (µm+µm/mm measuring radius)
|
0,02 + 0,0001
|
0,02 + 0,0001
|
0,02 + 0,0001
|
Axial runout deviation (µm+µm/mm measuring radius)
|
0,04 + 0,0002
|
0,04 + 0,0002
|
0,04 + 0,0002
|
Centering and tilting table
|
manual / automatic
|
manual / automatic
|
automatic
|
Table diameter (mm)
|
285
|
285
|
285
|
Table load capacity, centered (N)
|
600
|
600
|
400
|
Speed (rpm) 50 Hz / 60 Hz
|
0.2-15
|
0.2-15
|
0.2-15
|
Measuring path, motorized Z (mm)
|
350
|
500
|
900
|
Straightness deviation / 100 mm measuring path (µm), Z axis
|
0.15
|
0.15
|
0.15
|
Straightness deviation / total measuring path (µm), Z axis
|
0.3
|
0.4
|
0.9
|
Parallelism deviation Z-/C axis in tracing direction, measuring path (µm)
|
0.5
|
0.8
|
2
|
Measuring speed (mm/s), Z axis
|
0,1-30
|
0,1-30
|
0,1-30
|
Positioning speed (mm/s), Z axis
|
0,5 -100
|
0,5 -100
|
0,5 -100
|
Measuring path, motorized X (mm)
|
180
|
280
|
280
|
Straightness deviation /av. 100 mm measuring path (µm), X axis
|
0.4
|
0.5
|
0.5
|
Straightness deviation / total measuring path (µm), X axis
|
0.8
|
1.5
|
1.5
|
Perpendicularity X/C axis, measuring path (µm)
|
1
|
2
|
2
|
Positioning speed (mm/s), X axis
|
0,5-30
|
0,5-30
|
0,5-30
|
Measuring speed (mm/s), X axis
|
0,5-10
|
0,5-10
|
0,5-10
|